CAUTION: DISCONNECT LS200 FROM POWER SOURCE BEFORE SERVICING.
D IEC CONNECTOR R & R
D 1. REMOVE BACK HOUSING.
DISCONNECT LS200 FROM MAINS. IF MOUNTED ON MOBILE STAND,
DETACH TRANSFORMER ASSEMBLY FROM POLE. REMOVE FOUR
SCREWS AND REMOVE BACK HOUSING.
D 2. REMOVE IEC CONNECTOR.
DISCONNECT THREE WIRES FROM THE TERMINALS ON THE BACK OF
THE IEC CONNECTOR (POWER RECEPTACLE). LIFT IEC CONNECTOR
OUT OF THE FRONT HOUSING. RE-USE THE OLD FUSE TRAY IF IT IS NOT
DAMAGED. NEW FUSES OF THE CORRECT RATING SHOULD BE USED.
D 3. INSTALL REPLACEMENT IEC CONNECTOR.
ALIGN REPLACEMENT POWER RECEPTACLE WITH THE GROUND
(CENTER) TERMINAL AT THE BOTTOM OF THE HOLE IN THE CHASSIS.
PRESS RECEPTACLE INTO HOUSING CUTOUT. CONNECT WIRES PER
ASSEMBLY DRAWING. CONNECT THE BLACK WIRES TO THE IEC
CONNECTOR AT FI AND F2. THE CENTER TERMINAL IS THE GROUND
TERMINAL...CONNECT THE GREEN WITH YELLOW STRIPE (GROUND
WIRE) TO THE GROUND TERMINAL.
D 4. ATTACH BACK HOUSING OF POWER SOURCE.
MAKE SURE THE STRAIN RELIEF AND IEC CONNECTOR ARE IN THE
NOTCHES IN THE FRONT HOUSING. BE CAREFUL NOT TO PINCH
WIRES ON SECONDARY SIDE WHILE CLOSING HOUSING. POSITION
THE COVER ONTO THE POWER SUPPLY HOUSING AND FASTEN IT
WITH FOUR SCREWS.
D 5. TEST TO VERIFY SAFE OPERATION.
CONNECT LS200 TO POWER SOURCE AND VERIFY THAT ALL THREE
LAMPS ARE LIT. CONDUCT HI-POT TEST PER A-02861.
T E C H N I C I A N N O T E S :
____________________________________________________________________________
____________________________________________________________________________
____________________________________________________________________________
____________________________________________________________________________
16