CAUTION: DISCONNECT LS200 FROM POWER SOURCE BEFORE SERVICING.
G LAMP HOLDER R & R
G 1. REMOVE BEZEL.
(ALSO REFER TO PG.8 OPERATING INSTRUCTIONS PN441118-1ML)
PUT POWER SWITCH IN OFF POSITION AND REMOVE POWER CORD
CONNECTOR FROM A.C. POWER SOURCE. LAY UNIT DOWN WITH
HANDLE UP. UNSCREW HANDLE COUNTER CLOCKWISE. REMOVE
NUT, TETHER AND BEZEL ASSEMBLY.
DANGER! D O N O T T O U C H H O T L A M P S O R H O T L A M P
H O L D E R S ! (GLASS AND METAL COMPONENTS RETAIN HEAT
EVEN AFTER LAMPS ARE OFF SEVERAL MINUTES.)
G 2. REMOVE COOL LAMP(S).
AFTER A COOLING DOWN PERIOD OF AT LEAST 20 MINUTES,
LOCATE AND PULL THE LAMP HOLDING LEVER TO RAISE THE BASE
OF THE LAMP OUT OF ENGAGEMENT WITH THE LAMP SOCKET. PULL
THE LAMP HOLDING LEVER TOWARD THE BACK OF THE UNIT UNTIL
LEVER IS ALMOST VERTICAL. REMOVE EXPIRED LAMP.
G 3. REMOVE OLD LAMP HOLDER FROM LAMP HOLDER.
UNSCREW THREE SCREWS. CUT THE LAMP HOLDER WIRES AS
CLOSE TO THE LAMP HOLDER AS POSSIBLE. LIFT OLD LAMP
HOLDER OUT OF LAMP HOUSING.
G 4. ATTACH REPLACEMENT LAMP HOLDER TO LAMP HOUSING.
CONNECT REPLACEMENT LAMP HOLDER WIRES TO OLD WIRES
USING BUTT SPLICES PN 488377. KEEPING WIRES OUT FROM UNDER
THE LAMP HOLDER, POSITION LAMP HOLDER OVER THREE
THREADED BOSSES IN LAMP HOUSING. MAKE SURE THAT THE
WIRES DO NOT INTERFERE WITH THE LAMP CHANGING LEVERS.
INSERT AND TIGHTEN THREE SCREWS. INSTALL LAMP.
G 5. INSTALL BEZEL ASSEMBLY.
FOLLOW INSTRUCTIONS IN SECTION B-4 PG 14 TO ATTACH THE
BEZEL ASSEMBLY TO THE LS200.
G 6. TEST TO VERIFY SAFE OPERATION.
CONNECT TO A.C. POWER SOURCE AND ILLUMINATE LAMP. LET THE
UNIT STAY ON FOR FIVE MINUTES TO VERIFY PROPER OPERATION.
T E C H N I C I A N N O T E S :
____________________________________________________________________________
____________________________________________________________________________
____________________________________________________________________________
19