Directions for use Understanding the LCD messages 25
Block
The test cannot continue since the measured admittance is less than 0.2 mmho.
Possible causes:
Probe tip is lodged against canal wall
Ear canal occluded
Collapsed ear canal
Extremely small ear
Probe tip is clogged with cerumen
Possible solutions:
Reposition the probe tip
Perform otoscopy to check for occlusion
Remove cerumen from probe tip
Leak
The test cannot proceed since desired pressures within the ear have not been achieved.
Possible causes:
Probe tip is not completely sealed in the ear canal
Excessive movement of patient or practitioner
Probe tip dislodged during a test
-400 -200 0 200
2
0.5
0
1.0
1
GR
A
GR
C
PRESSURE - daPa
Ya - mmho
+200 Vea - cc