Directions for use Printing 226-Hz results 45
Change from manual to automatic printout
Automatic vs. Manual Printout (Switch #1)
Change printout format
Use Switches #2 and #3, located on the bottom of the Printer/Charger, to change printout
format. Printout options are shown below.
Auto Print Depress the ON portion of the switch to select this option.
This causes the printout to begin automatically once the
Handle is placed in the well, and data transfer is complete.
Manual Print
Depress the OFF portion of the switch to select this option.
This causes the printout to begin only when the PRINT button
is depressed.
1234
OFF
1234
OFF
Note
In the manual mode, a beep will occur as a reminder that data has been
transmitted; however, it is not necessary to wait for the beep before pressing
the PRINT button.
1234
OFF
1234
OFF
1234
OFF
OFF
1234
Tympanogram Only
(Switch 3)
Tympanogram and
Data (Switch 3)
Messages
(Switch 2)
No messages
(Switch 2)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Welch Allyn MicroTymp 3 and is the answer not in the manual?

Welch Allyn MicroTymp 3 Specifications

General IconGeneral
BrandWelch Allyn
ModelMicroTymp 3
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals