Operation with OxiTop
®
-IDS measuring heads Multi 3630/3620 IDS
26 ba77045d02 03/2020
Completing the
measurement
When the measurement is finished, the status LED flashes green for approx.
24 hours and the OxiTop
®
-IDS measuring head can be reached via radio.
5.2 Displaying and saving the momentary value
This function enables you to display the momentary pressure value of an active
sample and to save up to 10 momentary values per sample during the runtime
of the sample.
When 10 momentary values have been saved, no further momentary value can
be displayed. In this case, the <STO> key is without function.
The saved momentary values are output as a block at the end of the output to
the interfaces USB-A and USB-B or at the end of the csv file.
Displaying the
momentary value
If there is a radio link with the measuring head (LED magenta), you
can transfer and display the available measured values while a
measurement is running.
23
Using <M>, switch to the OxiTop set management.
24
Using <><>, select a set and open it with <MENU/ENTER>.
The data of the measuring heads are read out.
The reading out takes approx 5 seconds per OxiTop
®
-IDS measuring
head, with many measurement data it may take longer.
In the OxiTop set management, the set is given a checkmark [
] in the
status field.
25
If necessary, edit the evaluation of the sample
(e.g. if an OxiTop
®
-IDS measuring head is missing, or if there are obvi-
ously wrong values for an OxiTop
®
-IDS measuring head, see section
5.3 E
VALUATING THE SAMPLES).
26
Using <M>, switch to the OxiTop set management.
27
Using <><>, select the finished and read-out set.
28
Using <><>, select the finished and read-out set and, if necessary,
delete it with <F2>/[Delete].
The measuring heads are released for new measurements.
1
Using <><>, in the OxiTop set management highlight a Set with
running pressure measurement (Measuring mode Pressure p).
2
Open the details of the selected set with <MENU/ENTER>.
The measuring heads automatically call up the current data of the set.
The reading out takes approx 5 seconds per OxiTop
®
-IDS measuring
head, with many measurement data it may take longer.
3
Select a sample of the set.
4
Select a measuring head of the sample.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Xylem WTW OxiTop-IDS Multi 3630 IDS and is the answer not in the manual?

Xylem WTW OxiTop-IDS Multi 3630 IDS Specifications

General IconGeneral
BrandXylem
ModelWTW OxiTop-IDS Multi 3630 IDS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals