Ziehm Imaging, Inc. Vision Installation Manual MAN 06-0015, Rev. A Page 26
Maintenance Report: Tube Head Adjustment
Tube Head
Adjustment
Mechanical Pre-alignment Tube head Adjustment
Follow these instructions to install a replacement tube head or if tube head needs
alignment.
1. Turn off the Ziehm Vision and disconnect the power cord from the wall power
outlet
2. Remove tube head cover.
3. Set the C-Profile to zero using the rotation lock and the orbital locks to level the
C-Profile in both directions See Number 3 in the figure 15 a below.
4. Check the level of the C-Profile by placing a water level on the side of the C-Pro-
file.(Number 3 in the figure 15a below)
5. Place a level on the bottom of the image intensifier (Number 1 and 2 in the figure
15a below) if necessary adjust the I.I by loosening the mounting screws at the top
of the C-profile. If you can not level number 1 in figure below, insert a 1mm steel
shim to level the I.I.
Fig. 15 Placement of water levels
6. Check the tilt adjustment of the generator mounting plate. This must be level from
side to side so the generator can be adjusted properly. Loosen the two rear mount-
ing bolts to adjust the mounting plate. See number 5 in figure 15.
7. Check the horizontal tilt of the generator by placing the water level (Number 4 in
figure 15 on the top of the generator tube port. Adjust by loosening the mounting
bolts in figure 15 and tilting the generator until the generator is level.
Shim