EasyManuals Logo
Home>Agilent Technologies>Switch>34980A

Agilent Technologies 34980A Service Guide

Agilent Technologies 34980A
164 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #67 background imageLoading...
Page #67 background image
Calibration Procedures 3
34980A Service Guide 51
Calibration Process
The following general procedure is the recommended method to complete a full
instrument calibration.
1 Read “DMM Test Considerations" on page 55 and “Plug-in Module Test
Considerations" on page 67.
2 Perform the verification tests to characterize the instrument (incoming data).
3 Unsecure the instrument for calibration (“Calibration Security" on page 49).
4 Perform the DMM adjustment procedures (“Internal DMM Adjustments" on
page 62).
5 Perform the DAC adjustment procedures if either the 34951A (“34951A 4-Ch
Isolated DAC Module" on page 68) or 34952A (“34952A Multifunction Module" on
page 73) is installed.
6 Secure the instrument against calibration.
7 Note the new security code and calibration count in the instrument’s maintenance
records.
Aborting a Calibration in Progress
Sometimes it may be necessary to abort a calibration after the procedure has already
been initiated. You can abort a calibration at any time on any module by turning off the
power. You can abort a calibration on the internal DMM or the 34951A 4-Ch Isolated
DAC by issuing a remote interface device clear message. You can abort a calibration on
the 34952A Multifunction Module by sending the
CALibration:ABORt command.
NOTE
The 34951A 4-Ch Isolated DAC is intended to be adjusted frequently to
compensate for changes in the module’s environment. These
adjustments can be made either volatile or non-volatile. See page 68 for
details.
CAUTION
If you abort a calibration in progress when the instrument is attempting
to write new calibration constants to EEPROM, you may lose all
calibration constants for the function. Typically, upon re-applying
power, the instrument will report error 705 Cal:Aborted. You may also
generate errors 740 through 746. If this occurs, you should not use the
instrument until a complete re-adjustment has been performed. For a list
of calibration error numbers, see page 86.
深圳德标仪器 135-1095-0799

Table of Contents

Other manuals for Agilent Technologies 34980A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34980A and is the answer not in the manual?

Agilent Technologies 34980A Specifications

General IconGeneral
BrandAgilent Technologies
Model34980A
CategorySwitch
LanguageEnglish

Related product manuals