Calibration Procedures 3
34980A Service Guide 61
Optional AC Performance Verification Tests
These tests are not intended to be performed with every calibration. They are provided as
an aid for verifying additional instrument specifications. There are no adjustments for
these tests; they are provided for performance verification only.
Configuration: AC Volts
CONFigure[:VOLTage]:AC
LF 3 HZ:SLOW
[SENSe:]VOLTage:AC:BANDwidth 3
1
Make sure you have read “DMM Test Considerations" on page 55.
2 Select the AC Volts function and the 3 Hz filter.
3 Select each range in the order shown below. Provide the input voltage and frequency
indicated. Compare measurement results to the appropriate test limits shown in the
table below. (Be certain to allow for appropriate source settling.)
Input Error from Nominal
Voltage Frequenc
y
Range 24 hour 90 day 1 year
1 V 20 Hz 1 V ± 700 µV ± 900 µV ± 1 mV
1 V 20 kHz 1 V ± 700 µV ± 900 µV ± 1 mV
1 V 100 kHz 1 V ± 6.3 mV ± 6.8 mV ± 6.8 mV
1 V 300 kHz 1 V ± 45 mV ± 45 mV ± 45 mV
10 V 1 kHz 10 V ± 7 mV ± 9 mV ± 10 mV
1 V 1 kHz 10 V ± 3.4 mV ± 4.5 mV ± 4.6 mV
100 mV 1 kHz 10 V ± 13 mV ± 14 mV ± 14 mV
深圳德标仪器 135-1095-0799

Table of Contents

Other manuals for Agilent Technologies 34980A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34980A and is the answer not in the manual?

Agilent Technologies 34980A Specifications

General IconGeneral
BrandAgilent Technologies
Model34980A
CategorySwitch
LanguageEnglish

Related product manuals