EasyManuals Logo
Home>Agilent Technologies>Switch>34980A

Agilent Technologies 34980A Service Guide

Agilent Technologies 34980A
164 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #88 background imageLoading...
Page #88 background image
72 34980A Service Guide
3 Calibration Procedures
Analog Output Adjustment
Install the 34951A module in the mainframe and allow a 1 hour warm-up before
performing these procedures.
This adjustment procedure sets a zero adjustment and a gain adjustment constant for
each DAC output. You must perform all the adjustments on one analog output channel
before adjusting the other analog output channel.
1 Install the module(s) in the instrument. Remove any inputs from the ABus connector.
2 Set the calibration security for the adjustment mode you desire, see page 68.
3 Send the CALibration:MODule? <slot> command to begin the auto-calibration
procedure for all channels on the module in
<slot>. You may use the
CALibration:MODule? ALL command to perform the auto-calibration on all
channels for all modules installed in the mainframe. The complete procedure may
take up to 1 minute per module. Be sure to set the remote command time-out value
appropriately.
4 The command returns a value of “+0” if the calibration is successful.
A returned value of “+1” indicates a calibration failure.
CAUTION
Remove any ABus connector before performing this procedure.
NOTE
Following the auto-calibration procedure, the DMM is left in its reset
state (DCV).
深圳德标仪器 135-1095-0799

Table of Contents

Other manuals for Agilent Technologies 34980A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34980A and is the answer not in the manual?

Agilent Technologies 34980A Specifications

General IconGeneral
BrandAgilent Technologies
Model34980A
CategorySwitch
LanguageEnglish

Related product manuals