Page #58 background image
2-10
Using the 86100C
Eye/Mask Mode
8 On the display, click the Trigger button to open the Trigger dialog box. Or, click
Setup > Trigger. Select the General Trigger Setup tab on the dialog box.
9 Locate the Bandwidth settings in the Trigger dialog box. Select Standard for
data rates up to 3.2 GHz. This is the default bandwidth setting. If the
instrument has the enhanced trigger installed (Option 001), select Divided for
direct or divided trigger rates between 2 GHz and 13 GHz.
10 Close the Trigger dialog box.
11 Press the front-panel Eye/Mask Mode button.
12 Click Setup > NRZ Eye Mode for a Non-Return-to-Zero data format or RZ Eye
Mode for a Return-to-Zero data format.
13 Press the front-panel Auto Scale button. The eye diagram should be displayed.