Press Test worklist connection
37
to verify that the information has been entered correctly
and the server connection can be established.
Import TLS certificate from server or USB flash drive:
You can use TLS for enhanced security when setting up PACS and Worklist server connections.
Enabling TLS requires a TLS certificate to be imported to the displaying unit from a server or
from a USB flash drive. If multiple TLS certificates are imported to the displaying unit, the
PACS/Worklist server will select the required TLS certificate automatically. For import from
server, ensure that the displaying unit is connected to a Wi-Fi or LAN network (see section 5.1.).
For import from USB flash drive, ensure that USB connection has been enabled for certificate
import and a USB flash drive connected to the displaying unit (see section 4.3.).
39
40
41
42
To import TLS certificates from a server:
Press Server import
38
.
Press the input field next to Certificate file
name
39
and enter the file name.
Press the input field next to Host name
40
,
and enter the hostname.
Press the input field next to Port number
41
and enter the port number.
Press Import
42
.
To import TLS certificates from a USB flash drive:
Press USB import and wait while the displaying unit searches for TLS certificates on the USB
flash drive.
Select the required TLS certificate and press Import
43
.
Note: When the TLS certificate has been imported, the name of the certificate file is shown below Imported
TLS certificates.
Server information Explanation
PACS name The name entered here is the one shown on the list of
available servers in the Export menu.
PACS AE Title/
Worklist server AE title
The title entered here is used by the server to identify the
displaying unit's individual software application.
Host name/
Worklist server hostname
IP-address, MAC address or full web address for the server.
Port number/
Worklist server port number
Network port number for the server.
6. Operating the displaying unit
6.1. Preparation and inspection of the displaying unit before use
Before using the displaying unit for a patient procedure, follow the relevant steps below and
any other necessary steps according to your organization's procedures and requirements for
preparation and inspection of this type of device.
43
20

Table of Contents

Other manuals for Ambu aView 2 Advance

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ambu aView 2 Advance and is the answer not in the manual?

Ambu aView 2 Advance Specifications

General IconGeneral
BrandAmbu
ModelaView 2 Advance
CategoryMonitor
LanguageEnglish

Related product manuals