en
Problem Possible cause Recommended action
Low picture quality. Dirty/damp screen. Wipe the screen with a clean cloth as
described in chapter 9.
Image Adjustment
settings are not optimal.
Open the Image Adjustment menu
to adjust colour, contrast, sharpness
and brightness.
The distal end in the
visualization device is
dirty/damp.
Refer to the Instructions for Use of the
visualization device.
Displaying unit
cannot identify
a connected
USB device.
USB connector port
is damaged.
Move the USB device to another
USB connector port.
USB cable or external
USB device is damaged.
Connect a new USB cable or
a new USB device.
The displaying unit has
internal communication
problems.
Perform a Power OFF Cycle
(follow the procedure above this table).
The USB connection
may be disabled
Login as an Administrator, go to Settings –
Setup – General settings – enable USB
connection.
Failed data transfer
from displaying unit
to USB flash drive.
USB format is incorrect. The displaying unit supports data transfer
to USB flash drives formatted as exFAT,
EXT4, NTFS, BTRFS and FAT32.
15. Warranty and replacement
Ambu warrants that the displaying unit (as defined in section 2.1.) will conform to the
specifications described by Ambu and be free from defects in material and workmanship for a
period of one (1) year from the date of invoice.
Under this limited warranty, Ambu will be responsible only for either supplying authorized
spare parts or replacement of the displaying unit, as Ambu may decide in its sole discretion.
In case of replacement of spare parts, the customer is obligated to provide reasonable
assistance to Ambu, including, where relevant, by customer’s biomedical technicians pursuant
to instruction from Ambu.
Unless otherwise expressly agreed in writing, this warranty is the only warranty which applies to
the displaying unit, and Ambu expressly disclaims any other warranty, expressed or implied,
including any warranty of merchantability, suitability or fitness for a particular purpose.
The warranty applies only if it can be established that:
a) The displaying unit has not been disassembled, repaired, tampered with, altered, changed,
or modified by persons other than technical personnel (except with the prior written
consent of Ambu or in accordance with the instructions in the Spare Part Replacement
Manual); and
b) The defects or damage to the displaying unit does not result from abuse, incorrect use,
negligence, improper storage, inadequate maintenance or use of unauthorized accessories,
spare parts, consumables, or supplies.
In no event shall Ambu be liable for any indirect, incidental, consequential or special loss or
damages of any kind (including without limitation loss of profits or loss of use), whether or not
Ambu is or should be aware of the possibility of such potential loss or damage.
The warranty applies only to the original customer of Ambu and cannot be assigned or
otherwise transferred.
In order to avail itself of this limited warranty, if requested by Ambu, the customer must return
the displaying unit to Ambu (at its own expense and risk of shipment). In compliance with
applicable regulations, any displaying unit that has come into contact with potentially
infectious material must be decontaminated before being returned to Ambu under this limited
warranty (pursuant to the cleaning and disinfection procedures in chapter 9). Ambu is entitled
to reject a displaying unit which has not been duly decontaminated, in which case this limited
warranty shall not apply.
39