12. Disposal
At the end of product life, clean and disinfect the displaying unit (see chapter 10). Before
disposing of the displaying unit, it is advised to delete all sensitive data (see section 9.4.).
Then, dispose the displaying unit in accordance with local guidelines for electrical and
electronic waste.
13. Technical product specifications
13.1. Standards applied
The displaying unit function conforms with:
IEC 60601-1 Medical electrical equipment – Part 1: General requirements for basic safety and
essential performance.
IEC 60601-1-2 Medical electrical equipment – Part 1-2: General requirements for safety –
Collateral standard: Electromagnetic compatibility - Requirements for test.
IEC 60601-2-18 Medical electrical equipment – Part 2-18: Particular requirements for
the basic safety and essential performance of endoscopic equipment
The power supply conforms with:
IEC 60601-1 Medical electrical equipment – Part 1: General requirements for basic safety and
essential performance.
13.2. Specifications for aView 2 Advance
Display
Resolution:
Orientation:
Display type:
Start up time:
1920 x 1080 pixels
Landscape
12.8” colour TFT LCD
5 seconds
Connections
2 x USB connection:
Digital video out:
Wi-Fi:
LAN:
USB 3.0 Type A
HDMI and 3G-SDI (see note)
Video format: 1920 x 1080 p, 60 fps
Supports IEEE Standards 802.11ac/a/b/g/n
Ethernet RJ45 connector, 10/100/1000 Mbps
Note: For 3G-SDI interface, it is recommended to use a quality cable with better shielding (e.g. RG6 grade).
Memory
SSD hard drive capacity: 32 GB including Operating System
Mounting interface
Mounting interface standard: VESA MIS-D, 75 C, VESA FDMI compliant display,
Part D, with centre located mounting interface
Electric power
Power requirement:
Battery type:
Battery Operation:
Protection against electric shock:
19.0 VDC; 3.43 ADC
14.4 VDC 6500 mAh
Typical battery runtime of a new, fully charged
battery (the displaying unit turned ON and scope
connected) is 2.5 hours or more depending on the
connected scope (see note)
Internally powered
Note: For more information about battery time with different scopes connected, please contact your
local Ambu representative.
35

Table of Contents

Other manuals for Ambu aView 2 Advance

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ambu aView 2 Advance and is the answer not in the manual?

Ambu aView 2 Advance Specifications

General IconGeneral
BrandAmbu
ModelaView 2 Advance
CategoryMonitor
LanguageEnglish

Related product manuals