Accessing the Clinician Menu
Vercise™ DBS Remote Control 3 Directions for Use
92104389-01 17 of 27
When an impedance measurement is taken, impedances are assessed between a contact and the
case (monopolar), and between pairs of contacts (bipolar). A green square indicates that impedance is
within the acceptable range. A yellow dot with a question mark indicates that impedance is outside of
the acceptable range.
Monopolar impedance values can be displayed by pressing the Down Arrow button from the
Impedances screen.
2. Select Measure to check impedances again.
3. Select Back to go back to the Clinician Menu.
Reset Stimulator
You can use the Remote Control to reset the Stimulator.
From the Clinician Menu:
Select Reset Stimulator. The Reset Stimulator? screen displays.