32
Temperature
Temperature allows the user to calibrate or enter the slope for the infrared (temperature device). The user can also enter
the slope for the fiber optic, if applicable.
4. Using the “+/-” keys to highlight “Temperature.”
5. Press ENTER.
Infrared
Infrared allows the user to calibrate or enter the slope for the infrared temperature control. This is the first screen
To appear after “Temperature” has been selected from the “Calibration Setup” screen.
1. From the “Temperature Calibration” screen, use the “+/-” keys to select “Infrared.”
2. Press the ENTER key to enter the “IR Calibration” screen.
Slope
Slope is used to enter a known slope for the Infrared temperature control.
1. Use the “+/-” keys to highlight “Slope”
2. Press the ENTER key.
3. Use the “+/-” keys to highlight “Clr” to delete the current value.
4. Use the ARROW and “+/-” keys to highlight and select the proper number.
5. Continue highlighting the appropriate numbers and pressing the ENTER key to accept the values.
6. Once the slope has been entered, use the ARROW and “+/-” keys to highlight “OK.”
7. Press the ENTER key to complete the IR slope entry.