50
Remote Operation
SYST:ERR?”
Result was “30, COMMAND UNRECOGNIZED”, check for more errors.
Description: This error code is generated when the commanded string
does not contain any valid keyword (e.g. Route, System, *IDN?, *RST,
….) at all.
SYST:ERR?”
Result was “36, ID IS OUT OF RANGE”, check for more errors.
Description: A non existing switch ID has been commanded.
Eg. Sending Route:Switch11 8. When switch 11 does not exist.
SYST:ERR?”
Result was “50, UNABLE TO AQUIRE IP ADDRESS”, check for more errors.
Description: DHCP is enabled (ON), but the IP address could not be
acquired dynamically.
Eg. Ethernet connection cable is disconnected.
SYST:ERR?”
Result was “51, FAN STALL”, check for more errors.
Description: On models with fans equipped with sensors, one or more fans
have stalled (only when this alarm has been enabled at the factory).
Note: Not applicable for most models.
SYST:ERR?”
Result was “52, INTERNAL TEMPERATURE EXCEEDS THERSHOLD”, check
for more errors.
Description: On models with temperature sensors, one or more
temperature sensors has exceeded its alarm threshold (only when this
alarm has been enabled at the factory).
Note: Not applicable for most models.
SYST:ERR?”
Result was “53, POWER SUPPLY FAILURE”, check for more errors.
Description: On matrices with redundant power supplies it will indicate that
one of the power supplies does not operate.
E.g. Damaged power supply or blown fuse on the power supply.
Result was “0”NO ERROR””, No more errors, error queue is empty.