Instructions for use BiliLux 61
Reprocessing
11.4 Reprocessing list
11.5 Reprocessing procedures
11.5.1 Validated reprocessing procedures
At the time of product-specific validation, the following reprocessing procedures
showed good material compatibility and effectiveness:
The effectiveness of the listed reprocessing procedures has been validated by
independent laboratories that are certified to the standard ISO 17025.
11.5.2 Disinfectants
Use disinfectants that are nationally approved and are suitable for the particular
reprocessing procedure.
11.5.2.1 Surface disinfectants
At the time of the test, the surface disinfectants listed in the following table showed
good material compatibility. They can be used in addition to the surface
disinfectants listed in the section "Validated reprocessing procedures".
The manufacturers of the surface disinfectants have verified at least the following
spectra of activity:
– Bactericidal
– Yeasticidal
– Virucidal or virucidal against enveloped viruses
Observe the specifications of the surface disinfectant manufacturers.
Other surface disinfectants are used at one’s own risk.
Components Surface
disinfection
with cleaning
Manual
cleaning
followed by
disinfection by
immersion
Machine
cleaning with
thermal
disinfection
Steam
sterilization
Special
reprocessing
measures
Phototherapy
light
YesNoNoNoNo
Spring arm Yes No No No No
Trolley Yes No No No No
Radiometer Yes No No No No
Procedure Agent Manufacturer Concentration Contact time Temperature
Surface
disinfection with
cleaning
Dismozon
plus
BODE Chemie 1.6% 15 min —