Instructions for use BiliLux 67
Service
– There is no visible damage to device or power cables
4 Observe the instructions for use and check that all components and accessories
needed to use the product are present.
5 Check the electrical safety in accordance with the IEC 62353 or IEC 60601
standard.
12.5 Maintenance
The following table shows the preventive maintenance intervals. This table is not
related to warranty periods and no warranty is implied by the intervals listed.
The BiliLux LED panel does not need to be exchanged during the expected service
life of the device.
CAUTION
Risk of injury
The use of LED lights that are not approved by the manufacturer can affect the
safety and effectiveness of the phototherapy.
► Use only LED lights recommended by Dräger. Contact DrägerService for further
information.
12.6 Maintenance instructions
Following any maintenance procedure, perform the functional checkout procedure
on page 38.
12.6.1 Exchanging the clock battery
1 Remove the back cover from the phototherapy light.
2 Remove the old clock battery.
3 Install the new clock battery (CR2032).
4 Reinstall the back cover on the phototherapy light.
5 Set the date and time.
12.6.2 Calibrating the radiometer
Calibration must be performed by specialized service personnel. Package the
radiometer in the protective case it was shipped in, and send it to DrägerService.
Component Interval Measure Target group
Clock battery Every 6 years Exchange Service personnel
Radiometer Every 2 years Calibrate Specialized service
personnel
Spring arm joints Every 1 year, or as
required
1)
1) When the spring arm no longer maintains its position
Tighten Users or service
personnel