EasyManuals Logo
Home>Dräger>Medical Equipment>Oxylog 2000

Dräger Oxylog 2000 Instructions For Use

Dräger Oxylog 2000
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #19 background imageLoading...
Page #19 background image
Wipe disinfecting
Ventilator and medical gas hose
Follow the manufacturer's instructions.
Remove heavy soiling with a disposable cloth first.
Do not allow any liquid to enter the ventilator or
medical gas hose!
Bath disinfecting
Disassembled parts of the breathing valve, flow sensor,
ventilation hose and flow measuring hoses should be
Follow the manufacturer's instructions.
Agitate parts thoroughly in the solution.
Do not clean with a hard brush!
Do not allow any objects to enter the breathing valve
or flow sensor!
Allow to dry completely.
The breathing valve and flow measuring hoses may
not function correctly if water remains in these parts.
Sterilizing
Sterilize if necessary.
The disassembled parts of the breathing valve, the flow
sensor, the angled connector, the flow measuring hoses
and the ventilation hose
can be sterilized for at least 3 minutes in hot steam
at 134 °C, according to EN 285 (sterilisation, steam
sterilisation, major sterilisation).
Sterilisation for longer than 18 minutes is permissible,
but shortens the service life of the accessory.
Only reuse accessory if
it is undamaged and
the readiness for operation check has been
successfully completed.
Observe accessory service life
The parts of the ventilation valve, flow sensor, angled
connector, measuring hoses and ventilation hose are
resistant to the recommended disinfectants and are heat-
resistant during sterilisation.
However, wear occurs each time the parts are
disinfected and sterilized. After conditioning, the parts
should therefore be examined for cracking and
permanent deformation.
Damaged or deformed parts must be replaced.
After care
Reassemble, page 20.
Connect to power supply, page 21
and gas supply, page 25.
Check readiness for operation, page 28.
19
Care
Cleaning and disinfecting
Sterilizing
Rinse parts thoroughly with distilled water.
Disinfectant residues can cause the rubber disc
to become jammed in the breathing valve.

Other manuals for Dräger Oxylog 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 and is the answer not in the manual?

Dräger Oxylog 2000 Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals