Operation
Use a machine which has been cleaned and
disinfected and is ready for operation.
Care, page 17
Preparation, page 20
Checking readiness for operation, page 28
In order to detect any malfunctions quickly, the following
parameters should be monitored during ventilation:
Check airway pressure, e.g. the PEAK and PEEP
parameters, on the pressure gauge or on the display
Check minute volume (set by means of ”»Freq.« and
»VT«”) on the display.
Using controlled ventilation IPPV
For ventilation frequencies of 5 to 40 1/min.
During IPPV, the minimum ventilation frequency is limited
to 5 1/min by Oxylog 2000.
1 Switch for ventilation modes set to IPPV.
To speed up settings, set the scale range to the same
colour on the »Freq.« and »VT« knobs. This ensures that
the ventilation parameters are set as appropriate to the
patient group concerned, namely infants / children /
adults.
2 Set »Freq.« and »VT« knobs.
7
Operation
Using controlled ventilation IPPV
22
1
Body weight Freq. VT
(kg) 1/min litres
Green range 30 to 40 0.1 to 0.3
for infants
(7.5 to 20)
Blue range 20 to 30 0.3 to 0.8
for children
(20 to 40)
Brown range 5 to 20 0.8 to 1.5
for adults
(over 40)
00129033

Other manuals for Dräger Oxylog 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 and is the answer not in the manual?

Dräger Oxylog 2000 Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals