Positioning Oxylog 2000
For stationary use:
● Place on a level surface where it cannot slide or fall
or:
● Hang from the headboard of a bed.
or:
● Hang from a wall rail.
For mobile use in vehicles:
● Hang Oxylog 2000 in ventilator holder.
1 Hang ventilator from the bar on the holder using
clamps.
2 Swing ventilator upwards until it engages.
To remove the ventilator:
3 Push the release catch upwards.
Connecting gas supply
Take care when handling O2 !
Secure O2 cylinders so they cannot fall over and keep
away from excessive heat.
Do not grease or lubricate O2 fittings, such as cylinder
valves and pressure reducers, and do not handle with
greasy hands – fire risk!
Only open / close cylinder valves by hand and rotate
smoothly. Do not use tools.
O2 makes all fires burn more fiercely!
No smoking and no naked lights.
25