Checking ventilation
1 Set »VT« knob to 0.5 L.
2 Set »Freq.« knob to 12 1/min.
3 Set »TI : TE« knob to 1 : 2.
4 Set »Pmax« knob to 60 mbar.
5 Set »PEEP« knob to 5 mbar.
6 Set switch for ventilation modes to IPPV.
7 Set ON/OFF switch to I (ON).
● The ventilator carries out an electrical and pneumatic
self test.
● Display:
The software version is shown in the bottom line.
● The red alarm indicator and the display illumination
light up briefly.
● The alarm tone (beep) sounds twice. For safety
reasons, the alarm tone is output on two mutually
independent channels. Both are tested in the self test.
This is why the alarm tone sounds twice for the same
length of time.
● The green LED "DC power available" lights up
constantly when an external power supply in the range
11 to 13 V DC is connected.
The self test is complete within no more than 6 seconds.
● Display:
● Oxylog 2000 ventilates the test lung with the set
ventilation pattern.
29
Checking readiness for operation
Checking ventilation
Self test O.K.
Self test
SW-version xx.xx
1
2
D
Oxylog 2000
345
6
7