5.3 Warnings
Error
524332 -
self-test
failed
The MicroTubes intake/discharge
could not be correctly moved during
the self-test
Make sure that the wells/doors
for MicroTubes are not blocked;
re-start the device and carry out
the process once again
Error
524333 -
self-test
failed
The pneumatics system could not
generate the necessary mass flow
during the self-test
make sure that the gas outlet is
not sealed (e.g. by fingers, con-
tamination)
Error
524336 -
self-test
failed
The camera could not be correctly
accessed during the self-test
Repeat the self-test
Error
2097154 -
self-test
failed
The component driver could not be
correctly initialized during the self-
test
Repeat the self-test
Error Cause Remedy
Warnings Cause Remedy
X-act 7000 - low battery The battery charge of the
X-act 7000 is low
Promptly insert new bat-
teries in the X-act 7000
(see IFU)
Measurement - aborted User has manually
aborted the measurement
Start a new measurement
if necessary
Service - expired Service interval exceeded Servicing of the X-act
7000 by DrägerService is
recommended
MicroTubes - used up All 10 measurements of
the inserted MicroTubes
have been used up
Use new MicroTubes for
further measurements
MicroTubes - not found The RFID tag of the Micro-
Tubes was recognized, but
the MicroTubes could not
be drawn in
Check whether the Micro-
Tubes were inserted far
enough into the X-act
7000 for them to be cap-
tured by the automatic
intake; if necessary, push
the MicroTubes in a bit
faster and with slightly
more pressure; if no mea-
surement is intended,
remove the MicroTubes
from the immediate vicinity
of the X-act 7000

Table of Contents

Other manuals for Dräger X-act 7000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger X-act 7000 and is the answer not in the manual?

Dräger X-act 7000 Specifications

General IconGeneral
BrandDräger
ModelX-act 7000
CategoryTest Equipment
LanguageEnglish

Related product manuals