16.4 Steam sterilising
Individual parts can be sterilised either un-
packed or packed in a container or in sterile
product packaging� Use suitable sterile product
packaging�
Warning
Incorrect sterilisation inhibits effec-
tiveness and can damage the
product
• Only steam sterilisation is permitted�
• Do not exceed 134 °C�
• Observe the process parameters
• Do not use any other procedures
• Place filter paper on the sterilisation
tray
• Always sterilise the parts for sterilisa-
tion together with instruments, as this
ensures even temperature
distribution�
• Avoid any metal contact of the parts
for sterilisation with either any other
instruments or with the sides of the
tray
• Place parts for sterilisation away from
the heating rods and on the middle
rack so there is no contact with the
inside walls of the system�
• The product must be sterilised in a
disassembled state only�
Temperature: 134 °C
Pressure above the ambient surround-
ings: 2�16 bar / 0�216 MPa
Working time: 5 min
• Parts for sterilisation must be placed in a small
steam steriliser before every use in accord-
ance with EN 13060 and with sterilisation cy-
cle Type B with subsequent drying�
• Carefully observe the manufacturer‘s instruc-
tions regarding usage of the steam steriliser
and the correct placement of parts to be
sterilised�
• Observe all national and local guidelines and
standards regarding the sterilisation of medi-
cal products
33
34

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Durr Dental VC 45 and is the answer not in the manual?

Durr Dental VC 45 Specifications

General IconGeneral
BrandDurr Dental
ModelVC 45
CategoryDental equipment
LanguageEnglish

Related product manuals