Installation
7. Set-up
7.1 Room for set-up
• Medical treatment rooms or operating rooms�
• Room temperature must not fall below +5 °C
in winter and must not exceed +40 °C in
summer�
• If the appliance is first connected during win-
ter months, allow sufficient time after transport
and unpacking for the unit to warm to room
temperature before initial operation� (Danger
of short circuit due to build up of
condensation�)
7.2 Set-up alternatives
• Set up the VC 45 on a stable flat and raised
surface next to the treatment unit�
Operation of the VC 45 directly on the
floor is not permitted
• Set up the VC on the special cart (available as
special accessory)
2
8. Electrical Connection
Before first use (during commissioning), check
that the supply voltage is the same as that given
on the model identification plate�
• Connect to the power supply using the low
power socket (10) on the underside of the VC
45 and a plug for low power units�
Only use a cable with an angled plug
connection�
10