Platinous User’s Manual 97
Troubleshooting
8.3 User-level Servicing
8.3.1 Replacing Fuses
When a fuse blows, replace it with one of the included fuses.
Note: If a new fuse blows soon after being installed, contact the place of purchase or
Espec North America.
Procedure:
1. Set the main power switch in the OFF position.
2. Open the electric parts compartment door.
3. Replace the blown fuse with a new one.
4. Close the electric parts compartment door.
8.3.2 Changing the Chamber Lamp Bulb
Change the chamber lamp bulb after every 6,000 hours of service or whenever it blows. Use only the
included bulb. Replacement will require a Phillips screwdriver.
HOT! Shortly after the lamp goes out, it is HOT. Be careful of burns.
FRAGILE! Be careful when handling the chamber lamp bulb. It will
break if dropped or struck hard against objects.
Procedure:
1. Set the main power switch in the OFF position.
2. Remove the screw from the chamber lamp cover. Use a Phillips screwdriver.
FIGURE 8.49. HOW TO DETACH THE CHAMBER LAMP COVER