Platinous User’s Manual 59
Operation
6.2 Specimen Safety Device Setup
6.2.1 Overheat Protector
Always set the overheat protector BEFORE beginning tests. Set it
according to the type of specimens and test conditions.
Unless properly set, the overheat protector will not trip in the event of
chamber trouble, which could lead to specimen damage. For details on
the overheat protector setting, See “Specimen Protection (Safety
Devices)” on page 8.
Procedure:
1. Turn the dials until a suitable temperature is displayed.
FIGURE 6.38. OVERHEAT PROTECTOR
6.2.2 Absolute High/Low Limit and Upper Deviation Limit Temperature
(& Humidity) Alarms
Always set the absolute high/low limit temperature (& humidity)
alarms BEFORE beginning tests. Set alarms according to the type
of specimens and test conditions.
Unless properly set, the absolute high/low limit temperature (humidity)
alarms will not be triggered in the event of chamber trouble, which could
lead to specimen damage. For details on alarm settings, See “Specimen
Protection (Safety Devices)” on page 8.