Installation
48
Platinous User’s Manual
5.7.3 To Supply Water to the Humidifying Tray
To supply water to the humidifying tray, it is necessary to run the chamber. Water is supplied automatically
as long as humidity control is turned ON. The chamber is set to 23°C and humidity control of OFF when the
chamber is shipped from the factory. For details on chamber setup, see the SCP-220 Controller Manual.
Procedure:
1. Remove lower air grille. Set the main power breaker in the ON position.
2. Press the POWER key on the operating panel. The display will come on and the main menu will appear
shortly.
3. Go to monitor mode from the main menu. Press the MONITOR box.
4. If humidity control is ON, proceed to step 7. If OFF, press the MAIN MENU key and then the CON-
STANT SETUP box to get the Constant Setup screen.