The CRIT-LINE III System
1-8
____________________________________
1.7 Power Supply Adapter
The power supply adapter supplied with the CLM III must be used as the connector
to the AC power source. During normal monitoring, the power supply adapter
should be plugged into a grounded AC wall outlet and connected to the CLM III via
the power supply socket on the back panel (see Figure 2). (European and Japanese
Power Supplies are not earth grounded.) The CLM III has dual classification. It is a
class I device when operated on internal batteries and a class II device when
connected to an AC outlet
WARNING: Fresenius suggest that the brick portion of the power supply be
mounted to the top of the Crit-Line Monitor or to the dialysis machine using Velcro,
plastic ties, or other means in order to prevent the brick from “hanging” down from
the power jack which can result in the power jack coming loose or even breaking.
Mounting the brick portion of the power supply will also keep it off the floor and
less likely to get wet. See warning message below.
WARNING:
TO PREVENT POWER SUPPLY FROM OVERHEATING OR FAILURE:
KEEP POWER SUPPLY DRY
1. Do not place the power supply in or near an area that is or has the potential to
become wet.
2. Do not use the power supply that shows signs of corrosive build-up
3. Do not use a power supply that shows signs of damage such as frayed cords.
4. Protect the power supply from dropping.