Performance Characteristics
7-4
7.2 Standard Specifications
INSTRUMENT RANGE INPUT
Hematocrit (Hct): 10 Hct – 60 Hct 12 VDC/1 Ampere/12 watts
Oxygen Saturation: 55% – 100% 1.2A max
(Hct 18)
POWER SUPPLY UNIT
OPERATING AND STORAGE Universal regulated supply
TEMPERATURE 100VAC – 240VAC
50º F to 104º F 12 VDC/1.25 Ampere Output
(10º C to 40º C)
Avoid extreme temperatures during COMMUNICATIONS PORTS
transportation (<32º, >110 Fº) 8-position keyed modular jack
(Keyed RJ-45) serial port
HEMATOCRIT ACCURACY DB female parallel port
10 Hct – 60 Hct: ± 1 Hct SD
ELECTRICAL SHOCK PROTECTION
OXYGEN SATURATION ACCURACY Class II
55% – 100%: ± 2 Sat % Hct 18 Internally powered
Type BF
ACCESS BLOOD FLOW
Estimates flow rates WATER PROTECTION
50ml/min – 2500ml/min ± 15% Splash proof
RECIRCULATION CAPABILITY ANESTHETIC SUITABILITY
Estimates recirculation values>4% Not suitable
BATTERY CAPACITY MODE OF OPERATION
2 hours continuous on full charge Continuous
FULL CHARGE TIME BATTERY
36 hours Type: Rechargeable sealed lead
battery (12V, 2A-h) Battery is replaceable
PHYSICAL DIMENSIONS Polarity: ± (labeled on battery)
5.25” H, 8.25” W, 11.63” L, 5 lbs. Mode of insertion: Via 2 pin
Molex connector (replaceable)
HD BLOOD FLOW RATE
CAPABILITY
50 ml/min – 1300 ml/min
INTERNAL DATA STORAGE
Sufficient to store up to 26 hours of
data indefinitely