CIP / AxiChrom manifold
A CIP / AxiChrom manifold with four individually controlled valves enables UNICORN to
control CIP with up to four inlets and control processing together with a connected
AxiChrom column.
See Using the CIP/AxiChrom manifold option, on page 139 for more information about
connections for CIP. For Intelligent Packing with AxiChrom columns, see AxiChrom
manuals for details on connection.
Anybus™ X-gateway
The system can be provided with an optional Anybus X-gateway to enable communication
between ÄKTAprocess and the customer network. The signals transferred can both be
analog process readings, digital status and handshake signals. The Anybus X-gateway
is located inside the electrical cabinet. The interface to the gateway is PROFIBUS Slave
through a M12 connector in the bottom of the electrical cabinet.
The Anybus X-gateway copies I/O-data in both directions, thus enabling data exchange
between two optically isolated PROFIBUS networks. The Anybus X-gateway connections
can be used for many different applications. See the PROFIBUS Communication Interface
documenation for a description of how the different I/O-data is addressed in the mem-
ory space.
Contact your local GE Healthcare representative for more information.
ÄKTAprocess Operating Instructions 29-0252-49 AA 71
3 System description
3.4 Optional components

Table of Contents

Other manuals for GE AKTAprocess

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE AKTAprocess and is the answer not in the manual?

GE AKTAprocess Specifications

General IconGeneral
BrandGE
ModelAKTAprocess
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals