CARESCAPEPDM
DisinfectantManufacturerIngredientsTypeofdisinfectant
PDISuperSani-Cloth
Germicidal
DisposableWipes*
Professional
DisposablesInc.
n-Alkyl(68%C12,32%C14)
dimethylethylbenzylammonium
chlorides0.25%
n-Alkyl(60%C14,30%C16,5%
C12,5%C18)dimethylbenzyl
ammoniumchlorides0.25%
Isopropanol55.00%
Quaternary
ammoniumand
alcohol
PDIEasyScreen
CleaningWipes*
Professional
DisposablesInc.
Isopropylalcohol70%Alcohol
OxivirTBWipes*
SealedAirHydrogenperoxide0.1to1.5%
Benzylalcohol1to5%
Accelerated
hydrogenperoxide
Ethanol70to96%
Generic
EthanolAlcohol
Isopropylalcohol
max.60%byweight
Generic
IsopropylalcoholAlcohol
Bleach[dilutetomax.
of0.65%]
Generic
SodiumhypochloriteChlorine
*AvailableintheUnitedStates.
Disinfectingexternalsurfaces
Thedecisiontodisinfectmustbemadepertheinstitution’srequirementswithan
awarenessoftheeffectontheintegrityofthedevice.Donotuseexcessivedrying
techniques(e.g.,oven,forcedheat,sundrying).
1.Disconnectthedevicefromthehost.
2.Removeallcablesandthebattery.
3.Closethebatterydoor.
4.Dampenasoftlint-freecloth,orusedisinfectantwipeswithoneofthepermitted
disinfectants.
5.Wringexcessliquidfromtheclothandwipetheexteriorsurface.Payspecial
attentiontohard-to-cleanareas,likegroovesandcrevices.Allowdisinfecting
solutiontoremainondeviceperdisinfectantinstructionsorperhospital
guidelines.
Anycontactofdisinfectingsolutionswithmetalpartsmaycausecorrosion.
Donotdamageorbendconnectorpinswhencleaningordrying.
Donotletuidpoolaroundconnectionpins.Ifthishappens,blotdrywith
acottonswaborsoftcloth.
6.Wipeoffthedisinfectingsolutionwithaclean,lightlymoistenedcloth.
7.Drythoroughlywithadry,lint-freeclothandletairdryforatleast30minutes.
Dryingtimesmayvarybasedontheenvironmentalconditions.
8.Installthebatteryandclosethebatterydoor.
9.Reconnectallcables.
10.Reconnectthedevicetothehost.
16
Cleaninganddisinfecting
2110145-001

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE CARESCAPE MP100R and is the answer not in the manual?

GE CARESCAPE MP100R Specifications

General IconGeneral
BrandGE
ModelCARESCAPE MP100R
CategoryMedical Equipment
LanguageEnglish

Related product manuals