UnityNetworkIDInterfaceDevice
Cleaningcautions
CAUTION
Donotapplypressurizedairorgastoanyoutletortubing
connectedtothemonitor.Pressuremaydestroysensitive
elements.
Permitteddetergents
●Water
●EnzymaticorneutralpHdetergents
Cleaningpointstonote
●WarrantydoesnotcoveranydamagescausedbyusingotherthanGEapproved
substancesandmethods.
●GEmakesnoclaimsastotheefcacyofthelistedsubstancesormethodsto
controlinfection.
●Yourhospitalguidelinespermitting,allcleaninganddisinfectionactivitiescan
becarriedoutatthebedside.
●Donotletliquidpoolaroundconnectionpins.Ifthisshouldhappen,blotdrywitha
soft,lint-freecloth.
●Donotuseexcessivedryingtechniques,suchasoven,forcedheat,orsundrying.
●Neverconnectanydeviceorappliedparttoapatientuntilitisthoroughlydry.
●Ifyoudiscoveranysignsofdeteriorationordamageinthedevice,discontinueits
use.
Generalcleaninginstructions
Followtheseinstructionstocleanthedeviceandothernon-appliedpartsunlessthe
otherpartsprovidepart-specicinstructions.Non-appliedpartsrefertothosepartsof
thesystemthatarenotindirectcontactwiththepatient.
1.Turnoffthepowertotheequipment.
2.Disconnecttheequipmentfromthepowersupply.
3.Removeallcables.
4.Dampenasoftlint-freeclothwithoneofthepermitteddetergents.
5.Wringexcessliquidfromtheclothandwipetheexteriorsurface.Payspecial
attentiontohardtocleanareas,likegroovesandcrevices.
Anycontactofcleaningordisinfectingsolutionswithmetalpartsmaycause
corrosion.
Donotdamageorbendconnectorpinswhencleaningordrying.
6.Removeanysoilassoonaspossibleafterusebywipingthedevice.
Donotletuidpoolaroundconnectionpins.Ifthishappens,blotdrywithacotton
swaborsoftcloth.
7.Wipeoffthecleaningsolutionswithaclean,lightlymoistenedcloth.
8.Drythoroughlywithadry,lint-freeclothandletairdryforatleast30minutes.
Dryingtimesmayvarybasedontheenvironmentalconditions.
60
Cleaninganddisinfecting
2110145-001