CARESCAPEB850Monitor
Cleaningnon-appliedparts,generalinstructions
Followtheseinstructionstocleanthemonitor,modules,EEGheadbox,andother
non-appliedpartsunlessthereareseparatepart-specicinstructions.Non-applied
partsrefertothosepartsofthesystemthatarenotindirectcontactwiththepatient.
Beespeciallycarefulwhencleaningthedisplayscreen.Theyaremoresensitiveto
roughcleaningmethodsthanforexamplethemonitorhousing.
1.Turnoffthepowertotheequipment.
2.Disconnecttheequipmentfromthepowersupply.
3.Removeallcables.
4.Dampenasoftlint-freeclothwithoneofthepermitteddetergents.
5.Wringexcessliquidfromtheclothandwipetheexteriorsurface.Payspecial
attentiontohardtocleanareas,likegroovesandcrevices.
Anycontactofcleaningordisinfectingsolutionswithmetalpartsmaycause
corrosion.
Donotdamageorbendconnectorpinswhencleaningordrying.
6.Removeanysoilassoonaspossibleafterusebywipingthedevice.
Donotletuidpoolaroundconnectionpins.Ifthishappens,blotdrywithacotton
swaborsoftcloth.
7.Wipeoffthecleaningsolutionswithaclean,lightlymoistenedcloth.
8.Drythoroughlywithadry,lint-freeclothandletairdryforatleast30minutes.
Dryingtimesmayvarybasedontheenvironmentalconditions.
9.Visuallyensurethatthedeviceisclean.Ifvisiblesoilremains,repeatcleaning
procedureuntilthedeviceisthoroughlyclean.
10.Reconnecttheequipmenttothepowersupply.
11.Turnonthepowertotheequipment.
Otherdevices’cleaninginstructions
Fordetailsaboutcleaningtheaccessories,seetheinstructionsforuseinthe
accessorypackage.
Donotreusesingle-usedisposableaccessories.
Refertotheuserdocumentationprovidedwiththekeyboard,mouse,andbarcode
readerforinstructionsonhowtocleanthem.Alwaysconsideryourhospitalguidelines
aswell.
Disinfection
Disinfectionpointstonote
●Alwayscleanbeforedisinfecting.
●Alwaysdilutecleaninganddisinfectantagentsaccordingtotheirmanufacturer’s
instructions.Alwaysconsideryourhospitalguidelinesaswell.
●Useonlythepermittedsubstances.
2110145-001
Cleaninganddisinfecting
57