PatientMonitorB105/B125
DisinfectantManufacturerIngredientsTypeofdisinfectant
PDIEasyScreenCleaning
Wipes*
1
ProfessionalDisposables
Inc.
●Isopropylalcohol70%Alcohol
OxivirTBWipes*
SealedAir●Hydrogenperoxide0.1
to1.5%
●Benzylalcohol1to5%
Acceleratedhydrogen
peroxide
Ethanol70to96%
1
Generic
●EthanolAlcohol
Isopropylalcoholmax.
60%byweight
1
Generic
●IsopropylalcoholAlcohol
Bleach[dilutetomax.of
0.65%]
Generic
●SodiumhypochloriteChlorine
*AvailableintheUnitedStates.
1
NOTE:Donotusethisdisinfectantfortouchscreenpanels.
Disinfectingexternalsurfaces
Thedecisiontodisinfectmustbemadepertheinstitution’srequirementswithan
awarenessoftheeffectontheintegrityofthedevice.Donotuseexcessivedrying
techniques(e.g.,oven,forcedheat,sundrying).
1.Turnoffthepowertothedevice.
2.Disconnectthedevicefromthepowersupply.
3.RemovetherecorderandE-modulefrommonitor.
4.Removeallcablesandbattery.
5.Closebatterydoor.
6.Dampenasoftlint-freecloth,orusedisinfectantwipeswithoneofthepermitted
disinfectants.
7.Wringexcessliquidfromtheclothandwipetheexteriorsurface.Payspecial
attentiontohardtocleanareas,likegroovesandcrevices.Allowdisinfecting
solutiontoremainondeviceperdisinfectantinstructionsorperhospital
guidelines.
●Anycontactofdisinfectingsolutionswithmetalpartsmaycausecorrosion.
●Donotdamageorbendanyconnectorpinswhencleaningordrying.
●Donotletuidpoolaroundconnectionpins.Ifthishappens,blotdrywith
acottonswaborsoftcloth.
8.Wipeoffthedisinfectingsolutionwithaclean,lightlymoistenedcloth.
9.Drythoroughlywithadry,lint-freeclothandletairdryforatleast30minutes.
Dryingtimesmayvarybasedontheenvironmentalconditions.
10.Reconnectthedevicetothepowersupply.
11.Reconnectallcables.
12.Turnonthepowertothedevice.
42
Cleaninganddisinfecting
2110145-001