CARESCAPECentralStation
5.Wringexcessliquidfromtheclothandwipetheexteriorsurface.Payspecial
attentiontohard-to-cleanareas,likegroovesandcrevices.
●Removeanysoilassoonaspossibleafterusebywipingthedevice.
●Anycontactofcleaningsolutionswithmetalpartsmaycausecorrosion.
●Donotdamageorbendconnectorpinswhencleaningordrying.
●Donotletuidpoolaroundconnectionpins.Ifthishappens,blotdrywith
acottonswaborsoftcloth.
6.Wipeoffthecleaningsolutionswithaclean,lightlymoistenedcloth.
7.Visuallyensurethatthedeviceisclean.Ifvisiblesoilremains,repeatcleaning
procedureuntilthedeviceisthoroughlyclean.
8.Drythoroughlywithadry,lint-freeclothandletairdryforatleast30minutes.
Dryingtimesmayvarybasedontheenvironmentalconditions.
9.Reconnectthedevicetothepowersupply.
10.Reconnectallcables.
11.Turnonthepowertothedevice.
Othermanufacturercleaninginstructions
Fordetailsaboutcleaninganddisinfectioninformationfordevices,device
components,supplies,andaccessoriesmadebymanufacturersotherthanGE,see
theapplicableinstructionsforuseprovidedbythemanufacturer.
Disinfection
Disinfectionpointstonote
●Alwayscleanbeforedisinfecting.
●Alwaysdilutecleaninganddisinfectantagentsaccordingtotheirmanufacturer’s
instructions.Alwaysconsideryourhospitalguidelinesaswell.
●Useonlythepermittedsubstances.
●Yourhospitalguidelinespermitting,allcleaninganddisinfectionactivitiescan
becarriedoutatthebedside.
●Beforedryingasurfaceafterwipingit,waittheminimumtimerequiredaccording
tothesubstancemanufacturer’sinstructions.
●Visuallyensurethatnosubstanceresidueremainsonthedevice.
Permitteddisinfectants
Aboutthefollowingtable:
●Allthirdpartytrademarksarepropertyoftheirrespectiveowners.
●Trademarknamesandproductavailabilitymayvarybycountry.Consultthe
columnthatlistsingredientstodetermineifanequivalentdisinfectantisavailable
inyourcountry.