ApexProandCARESCAPETelemetry
DisinfectantManufacturerIngredientsTypeofdisinfectant
PDISuperSani-Cloth
Germicidal
DisposableWipes*
Professional
DisposablesInc.
●n-Alkyl(68%C12,32%C14)
dimethylethylbenzylammonium
chlorides0.25%
●n-Alkyl(60%C14,30%C16,5%
C12,5%C18)dimethylbenzyl
ammoniumchlorides0.25%
●Isopropanol55.00%
Quaternary
ammoniumand
alcohol
PDIEasyScreen
CleaningWipes*
Professional
DisposablesInc.
●Isopropylalcohol70%Alcohol
OxivirTBWipes*
SealedAir●Hydrogenperoxide0.1to1.5%
●Benzylalcohol1to5%
Accelerated
hydrogenperoxide
Ethanol70to96%
Generic
●EthanolAlcohol
Isopropylalcohol
max.60%byweight
Generic
●IsopropylalcoholAlcohol
Bleach[dilutetomax.
of0.65%]
Generic
●SodiumhypochloriteChlorine
*AvailableintheUnitedStates.
Disinfectingexternalsurfaces(telemetrymonitoring
devicesonly)
Thedecisiontodisinfectmustbemadepertheinstitution’srequirementswithan
awarenessoftheeffectontheintegrityofthedevice.Donotuseexcessivedrying
techniques(e.g.,oven,forcedheat,sundrying).
1.Removeallcablesandthebatteries.
2.Closethebatterydoor.
3.Dampenasoftlint-freecloth,orusedisinfectantwipeswithoneofthepermitted
disinfectants.
4.Wringexcessliquidfromtheclothandwipetheexteriorsurface.Payspecial
attentiontohard-to-cleanareas,likegroovesandcrevices.Allowdisinfecting
solutiontoremainondeviceperdisinfectantinstructionsorperhospital
guidelines.
●Anycontactofdisinfectingsolutionswithmetalpartsmaycausecorrosion.
●Donotdamageorbendconnectorpinswhencleaningordrying.
●Donotletuidpoolaroundconnectionpins.Ifthishappens,blotdrywith
acottonswaborsoftcloth.
5.Wipeoffthedisinfectingsolutionwithaclean,lightlymoistenedcloth.
6.Drythoroughlywithadry,lint-freeclothandletairdryforatleast30minutes.
Dryingtimesmayvarybasedontheenvironmentalconditions.
7.Installthebatteriesandclosethebatterydoor.
8.Reconnectallcables.
2110145-001
Cleaninganddisinfecting
29