PRN50M+
Cleaningpointstonote
●Warrantydoesnotcoveranydamagescausedbyusingsubstancesormethods.
●Manufacturermakesnoclaimsastotheefcacyofthelistedsubstancesor
methodstocontrolinfection.
●Alwaysdilutecleaninganddisinfectantagentsaccordingtotheirmanufacturer’s
instructions.Alwaysconsideryourhospitalguidelinesaswell.
●Yourhospitalguidelinespermitting,allcleaninganddisinfectionactivitiescan
becarriedoutatthebedside.
●Donotletliquidpoolaroundconnectionpins.Ifthisshouldhappen,blotdrywitha
soft,lint-freecloth.
●Donotuseexcessivedryingtechniques,suchasoven,forcedheat,orsundrying.
●Neverconnectanydeviceuntilitisthoroughlydry.
●Ifyoudiscoveranysignsofdeteriorationordamageinthedevice,discontinueits
use.
Generalcleaninginstructions
Followtheseinstructionstocleanthedevice.
1.Turnoffthepowertothedevice.
2.Disconnectthedevicefromthepowersupply.
3.Removeallcables.
4.Dampenasoftlint-freeclothwithoneofthepermitteddetergents.
5.Wringexcessliquidfromtheclothandwipetheexteriorsurface.Payspecial
attentiontohard-to-cleanareas,likegroovesandcrevices.
●Removeanysoilassoonaspossibleafterusebywipingthedevice.
●Anycontactofcleaningsolutionswithmetalpartsmaycausecorrosion.
●Donotdamageorbendconnectorpinswhencleaningordrying.
●Donotletuidpoolaroundconnectionpins.Ifthishappens,blotdrywith
acottonswaborsoftcloth.
6.Wipeoffthecleaningsolutionswithaclean,lightlymoistenedcloth.
7.Visuallyensurethatthedeviceisclean.Ifvisiblesoilremains,repeatcleaning
procedureuntilthedeviceisthoroughlyclean.
8.Drythoroughlywithadry,lint-freeclothandletairdryforatleast30minutes.
Dryingtimesmayvarybasedontheenvironmentalconditions.
9.Reconnectthedevicetothepowersupply.
10.Reconnectallcables.
11.Turnonthepowertothedevice.
Cleaninginternalcomponents
Onaregularbasis,qualiedservicepersonnelshouldopenthewriterandblowdust
fromtheunitwithcompressedair.
66
Cleaninganddisinfecting