Chapter 7. Specifications
56 DM5E Operating Manual
7.1 Instrument Specifications
Operating Principle: Ultrasonic, pulse echo measurement method
Measuring Range: Depends on probe and material,
0.60mm to 508mm (0.025in to 20.00in) in IP to 1st BW Measurement Mode,
2.00mm to 127.0mm (0.079in to 5.00in) in Dual-Multi Measurement Mode,
the coating thickness range shall be 0.3mm to 2.50mm (0in to 0.098in).
Measuring Resolution: 0.001 inch default (selectable 0.001, 0.01 inch)
0.01 mm default (selectable 0.01, 0.1 mm)
Material Velocity Range: 0.0200 to 0.7362 in/msec
0.508 to 18.699 mm/msec
Material Velocity Resolution: 0.0001 in/msec
1 m/s
Units: inch or millimeter
Calibration: One-Point Cal, On-Block and Off-Block, Two-Point Cal