EasyManuals Logo

GE HygroPro User Manual

GE HygroPro
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #44 background imageLoading...
Page #44 background image
Chapter 3: Service & Maintenance
Page 38 HygroPro User’s Manual
3.3.3 Removing the Transmitter from the System
Complete the following steps to remove the transmitter from the
installation site:
1. Refer to Figure 2 on page 8 and use a 1-1/8” wrench on the probe
hex nut to unthread the transmitter from the fitting on the sample
system or process line.
2. Record the dew point of the ambient air.
3. Disconnect the cable from the transmitter module.
3.3.4 Removing the Probe from the Transmitter
To remove the probe from the transmitter, refer to Figure 9 on the
next page and proceed as follows:
1. Loosen the four captive screws on the bottom of the transmitter.
2. Carefully remove the metal plate without touching the sensor.
3. Carefully pull the probe out of the transmitter.
4. Disconnect the probe cable by turning the locknut at the top of the
probe. Then, detach the sensor.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE HygroPro and is the answer not in the manual?

GE HygroPro Specifications

General IconGeneral
BrandGE
ModelHygroPro
CategoryTransmitter
LanguageEnglish

Related product manuals