EasyManua.ls Logo

GE MOA 280i User Manual

GE MOA 280i
211 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #149 background imageLoading...
Page #149 background image
GE Analytical Instruments ©2006 13-7 DLM 14291 Rev. A
Dilute to the mark with nitrate-free, deionized water.
Invert the flask several times until the solid dissolves.
Transfer the stock solution to air-tight storage container.
Calculate the concentration of the stock solution. For NaNO
3
, the
concentration is:
Conc. of Stock = (mgNaNO
2
/85 mg/mmol)/10 mL x 1000 mL/L
Use this stock 100 mM standard to prepare dilutions to construct a standard
curve for the NOA. The concentrations of the dilutions will depend on the
nitrate levels in the samples. Prepare dilutions that contain nitrate at
concentrations greater than the samples, concentrations less than the samples
and concentrations approximately the same as the samples. For most cell
culture work, nitrate concentrations will be in the nanomolar to micromolar
range. Analysis of standard solutions containing 10 nM, 50 nM, 100 nM, 1 µM, 5
µM, 10 µM, 50 µM and 100 µM will allow construction a calibration curve
suitable for most samples. For urine and other samples that contain high levels
of NO, the standard curve may be extended to include standards with nitrate
concentrations greater than 100 µM.
Preparation of Dilute Standards
The best way to prepare dilute standards is by serial dilution. The containers
for the dilute standards must be free of nitrate contamination (rinse with
nitrate-free water before use). Microfuge tubes (1.5 mL) or similar disposable
containers are suitable containers. To prepare the 1 mL of the dilute standards:
Arrange 7 or 8 tubes in a rack and add 900 µL of nitrate-free water to each
tube.
Set one tube aside and label the tube Water Blank.
Add 100 µL of the stock solution to the first tube, mix thoroughly and label
the tube 10 mM (Note: the actual concentration will depend on the

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the GE MOA 280i and is the answer not in the manual?

GE MOA 280i Specifications

General IconGeneral
BrandGE
ModelMOA 280i
CategoryMeasuring Instruments
LanguageEnglish

Summary

3. MENUS AND CONTROL OVERVIEW

Analysis

Manages the NOA's operational modes (Start, Stand-by, Stop).

Calibration

Process for calibrating the analyzer for accurate measurements.

Maintenance

Accesses functions for analyzer upkeep and service.

5. INSTALLATION AND SETUP: GAS-PHASE MEASUREMENTS

NOA Setup for Gas-Phase Measurements

Configuration settings for gas-phase measurements.

6. CALIBRATION

Zero Gas Calibration

Procedure for calibrating the NOA using a zero gas standard.

NO Calibration Gas

Procedure for calibrating the NOA with a known concentration of NO gas.

12. MEASUREMENT OF NITRIC OXIDE AND NITRITE IN LIQUID SAMPLES

13. MEASUREMENT OF NITRATE, NITRITE AND NITRIC OXIDE IN LIQUID SAMPLES

15. MAINTENANCE

Changing the Vacuum Pump Oil

Procedure for changing the vacuum pump oil.

Cleaning the Chemiluminescence Reaction Cell

Procedures for cleaning the reaction cell.

16. TROUBLESHOOTING

Errors

List of error conditions and their remedies.

Possible Errors and Remedies

Detailed descriptions of common errors and solutions.

No Power to NOA

Diagnosing and resolving issues with no power to the analyzer.

No Display

Troubleshooting steps for a blank display.

Related product manuals