GE Analytical Instruments ©2006 DLM 14291 Rev. A
3-6
C O N T R O L
M E T H O D
S E T U P
S T A T U S
Fi
g
ure 3-8: Control Menu.
A N A L Y S I S
S T O P
S T A N D - B Y
Fi
g
ure 3-7:
nal
sis Menu
NOA in Start Mode
.
When the NOA is in the start mode, selecting Stop will:
Display a confirmation screen “Are you sure? continue / stop”
If the Stop option is selected, turn off the PMT, and ozone generator.
Run the vacuum pump for 2 minutes to clear residual ozone from the
system.
Turn off the vacuum pump.
Main Menu Options
The other options from the Main Menu are used to configure the NOA for the
different applications, perform the calibration for gas-phase measurements,
view warnings and errors, and view and install consumables.
Control
The Control Menu is used to setup the NOA for liquid or gas-phase
measurements, set the output intervals and the sensitivity of the NOA. The
setup can be stored as a method to permit easy switching between
applications. The Control Menu also permits viewing of the cooler temperature