EasyManua.ls Logo

GE MOA 280i User Manual

GE MOA 280i
211 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #186 background imageLoading...
Page #186 background image
GE Analytical Instruments ©2006 16-2 DLM 14291 Rev. A
After viewing the error occurrence and help screen, press ENTER or CLEAR to
return to the Error Menu. After viewing an error, the asterisks in the Error Menu
are removed.
A list of errors that have been detected will remain in the error stack until the
error stack is cleared as described below.
Possible Errors and Remedies
E 01 – Setup Data Corrupted, Check Before Running
When the power to the NOA is first turned on, the program checks the data stored
in the battery-backed RAM. If there is a problem with this data, E 01 is reported.
This error may be generated when installing a new version of the firmware. In
most cases, continue running after getting this error, but if the error continues,
contact Sievers.
E 02 – Cell Pressure was Above the Limit
When the NOA is running, this error is reported if the cell pressure rises above 100
torr. Make sure that the main power switch on the vacuum pump is in the on
position, the power cord from the pump is securely plugged into the rear of the
NOA and the vacuum hose connections to the NOA and Hopcalite trap are secure.
Select the Analysis/Stand-by Mode to start the vacuum pump without getting this
error and monitor the cell pressure in the Control/Status menu to help locate any
leaks.
E 03 – Ozone Supply Pressure was Below the Limit
The NOA monitors the supply pressure of oxygen to the ozone generator and if the
pressure drops below 4 psi, E 03 is reported. This error will be reported when the
oxygen cylinder is empty or if there is no oxygen flow into the NOA. Make sure that
the oxygen tank contains gas, the main valve is turned on and the outlet pressure
on the regulator is set to > 6 psig. If an E size cylinder and flow controller are

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the GE MOA 280i and is the answer not in the manual?

GE MOA 280i Specifications

General IconGeneral
BrandGE
ModelMOA 280i
CategoryMeasuring Instruments
LanguageEnglish

Summary

3. MENUS AND CONTROL OVERVIEW

Analysis

Manages the NOA's operational modes (Start, Stand-by, Stop).

Calibration

Process for calibrating the analyzer for accurate measurements.

Maintenance

Accesses functions for analyzer upkeep and service.

5. INSTALLATION AND SETUP: GAS-PHASE MEASUREMENTS

NOA Setup for Gas-Phase Measurements

Configuration settings for gas-phase measurements.

6. CALIBRATION

Zero Gas Calibration

Procedure for calibrating the NOA using a zero gas standard.

NO Calibration Gas

Procedure for calibrating the NOA with a known concentration of NO gas.

12. MEASUREMENT OF NITRIC OXIDE AND NITRITE IN LIQUID SAMPLES

13. MEASUREMENT OF NITRATE, NITRITE AND NITRIC OXIDE IN LIQUID SAMPLES

15. MAINTENANCE

Changing the Vacuum Pump Oil

Procedure for changing the vacuum pump oil.

Cleaning the Chemiluminescence Reaction Cell

Procedures for cleaning the reaction cell.

16. TROUBLESHOOTING

Errors

List of error conditions and their remedies.

Possible Errors and Remedies

Detailed descriptions of common errors and solutions.

No Power to NOA

Diagnosing and resolving issues with no power to the analyzer.

No Display

Troubleshooting steps for a blank display.

Related product manuals