Use
Daily inspections before use
4
LUCEA 50-100
IFU 01741 EN 11
27 / 60
4 Use
4.1 Daily inspections before use
Fig.9: Integrity of the lightheads
Integrity of the lightheads, brake screw cap
and mounting screws
1. Check the lightheads for chipped paint, im-
pact marks, any other damage, loose cov-
ers, etc.
2. Check that the cap protecting the brake
screw is properly seated.
3. Check that all the mounting screws are
present.
4. If a problem is noted, contact technical
support.
Fig.10: DF fork plugs
Fork plugs (DF version only)
1. Check that the grey plugs are correctly in-
stalled.
2. If a problem is noted, contact technical
support.
Fig.11: Stability/drift
Stability and drift of the system
1. Operate the device, making several move-
ments in order to swivel the extension
arms, the spring arms and the lightheads.
Ø The entire system should move easily
and smoothly.
2. Place the system in various positions.
Ø The entire system should remain in the
selected position, without any drift.
3. If a problem is noted, contact technical
support.

Table of Contents

Other manuals for Getinge Lucea 100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Lucea 100 and is the answer not in the manual?

Getinge Lucea 100 Specifications

General IconGeneral
BrandGetinge
ModelLucea 100
CategoryMedical Equipment
LanguageEnglish

Related product manuals