Technical specifications
Other characteristics
9
LUCEA 50-100
IFU 01741 EN 11
55 / 60
9.4 Other characteristics
Protection against electrical shock Class I
Medical device classification for Europe, Canada, Korea, Japan, Brazil &
Australia
Class I
Medical device classification for USA, China & Taiwan Class II
Protection rating for the device as a whole IP20
Protection rating of the lightheads IP20
GMDN code 12282 / 36843
EMDN code Z12010701 /
Z12010702
CE marking year 2011
Tab.22: Specifications relating to standards and regulations
9.5 EMC declaration
CAUTION!
Risk of malfunction of the device
If the device is used in conjunction with other equipment, its operation and
performance may be affected.
Do not use the device alongside other equipment or stacked with other
equipment except after observing the normal operation of the device and the
other equipment.
CAUTION!
Risk of malfunction of the device
The use of hand-held RF communications equipment (including antenna
cables and external antennas) alongside the device or specified cables may
affect the operation and performance of the device.
Do not use hand-held RF communications equipment at within 30cm of the
device.
CAUTION!
Risk of malfunction of the device
The use of a high frequency generator (e.g. electrosurgical unit) adjacent to
the device may affect its operation and performance.
If anomalous operation is observed, adjust the position of the lightheads un-
til the interference ceases.
CAUTION!
Risk of malfunction of the device
The use of the device in an unsuitable environment may affect its operation
and performance.
Do not use this device except in a professional healthcare facility.