6 Troubleshooting
Electronics/Optics
Problem Likely cause Corrective action
The lighthead does not turn
on.
Power cut Contact your facility’s technical
services
Does not switch over to
backup
Contact the Getinge technical
department
Other reason Contact the Getinge technical
department
The lighthead does not turn
off.
Communication problem Contact the Getinge technical
department
An LED does not come on. The LED board is defective Contact the Getinge technical
department
The remote control does not
operate the light
Registration problem Re-register the remote control
Battery level low Replace the batteries
Tab.11: Troubleshooting
Mechanical components
Problem Likely cause Corrective action
The sterilisable handle does
not click into place correctly
Sterilisation parameters (tem-
perature, time) exceeded
Check the operation of the
handle and in particular the
locking mechanism (audible
click)
Its maximum service life has
expired or the handle is twis-
ted or bent.
Replace the handle
The lighthead drifts Suspension tube not vertical Contact the Getinge technical
department
Ceiling structure unstable Contact the Getinge technical
department
Locking screw incorrectly ad-
justed.
Contact the Getinge technical
department
Lighthead moves too easily or
is difficult to move.
Locking screw incorrectly ad-
justed.
Contact the Getinge technical
department
Other reason Contact the Getinge technical
department
Tab.12: Mechanical anomalies and malfunctions