4
Use
Positioning the light
32 / 60
LUCEA 50-100
IFU 01741 EN 11
4.4 Positioning the light
4.4.1 Installing/removing the sterilisable handle
WARNING!
Risk of infection
If the sterile handle is not in good condition, there is a risk that particles
could fall from it into the sterile environment.
After each sterilisation and before using a sterilisable handle again, check
that there are no cracks.
Fig.20: Installing the sterilisable handle
Installing a sterilisable handle on the light-
head
1. Inspect the handle and check for cracks or
soiling.
2. Fit the handle to the mount.
3. Rotate the handle until its rotation is
locked.
Ø The locking button pops out of its hous-
ing.
4. Check that the handle is firmly in place.
Ø The handle is now locked in place and
ready for use.
Fig.21: Removing the sterilisable handle
Removing a sterilisable handle from the
lighthead
1. Press the locking button.
2. Remove the handle.

Table of Contents

Other manuals for Getinge Lucea 100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Lucea 100 and is the answer not in the manual?

Getinge Lucea 100 Specifications

General IconGeneral
BrandGetinge
ModelLucea 100
CategoryMedical Equipment
LanguageEnglish

Related product manuals