9 Removing the transport securing device
It is imperative that the transport securing device be removed.
Keep the transport securing device in a safe place since it must be installed again before transporting the
device.
The device may only be transported with the transport securing device installed.
To protect the device from damage during transport, the motor is fixed in place.
This transport securing device must be removed when the device is put into operation.
(a)
(b) (b)
(a)
 Remove the two screws (a) and spacer sleeves (b).
The transport securing device is installed in the opposite order.
10 Initial operation
 A type B residual current circuit breaker must be used if the device is additionally protected with a residual
current circuit breaker in the building installation. When using a different type, the residual current circuit breaker
may either not switch off the unit if there is a fault on the unit, or it may switch off the unit even though there is no
fault on the unit.
 Remove the transport securing device from the housing floor; see the "Removing the transport securing device"
chapter.
 Position the centrifuge in a stable and level manner in a suitable place. During set-up, the required safety
margin of 300 mm around the centrifuge is to be kept according to EN / IEC 61010-2-020.
When the centrifuge is running, according to EN / IEC 61010-2-020, no persons, dangerous
substances or objects may be within the safety margin of 300 mm around the centrifuge.
 Ventilation openings may not be blocked.
A distance of 300 mm must be maintained from the ventilation slots and openings of the centrifuge.
 Check whether the mains voltage tallies with the statement on the type plate.
 Connect the centrifuge with the power cord to a standard mains socket. For connection ratings refer to Chapter
"Technical specifications".
 Switch on the mains switch (switch position "").
The following displays appear one after the other:
1. The centrifuge model
2. The type number and program version
3. The last used centrifugation data
If the lid is closed, the message "Open the lid" is displayed.
In this case, open the lid to display the centrifugation data.