15.3 Setting options
t/min Runtime. Adjustable from 1 - 99 min, in steps of 1 minute.
t/sec Runtime. Adjustable from 1 - 59 s, in steps of 1 second.
Continuous run "--:--". Set the parameters t/min and t/sec to zero.
RPM Speed
A number value from 200 rpm to the maximum speed of the rotor can be set.
Adjustable from 200 RPM to 10000 RPM in steps of 10, and above 10000 RPM, in steps of 100.
For the maximum rotor speed, see the chapter "Appendix, Rotors and accessories".
RCF Relative centrifugal force
A number value can be set which results in a speed between 200 rpm and the maximum rotor speed.
Adjustable up to 10000 in steps of 1, and above 10000, in steps of 10.
It is only possible to input the relative centrifugal force (RCF) if the RCF display (
RCF ) is
activated.
The relative centrifugal force (RCF) depends on the centrifuging radius (RAD). After entering the
RCF, check to make sure that the correct centrifuging radius has been set.
RAD/mm Centrifuging radius
Adjustable from 10 mm to 250 mm, in steps of 1 mm.
For the centrifuging radius, see the chapter "Appendix, Rotors and accessories".
It is only possible to input the centrifuging radius if the RCF display (
RCF ) is activated.
DEC Brake stage. fast = short run-out time, slow = long run-out time.