DATA
ENTRY
(cant)
Offset
Range
The
3314A's
DC
Offset
can be set
from
-5V
to +5V in 2
offset
ranges. To assert
Range Hold
from
the
front
panel, press the OFFSET key and then
the
BLUE shifted
DOWN ARROW key. To assert Range Hold
from
the HP-IB, send
"OFRHx"
where
x is
the
Range Number. Note
that
the
offset
will change
when
Range Hold is
asserted
from
the HP-IB
if
the current
entry
is outside
the
normal limits
of
the
selected range. Pressing the OFFSET key and then the BLUE shifted DOWN
ARROW key or
,.OFRHO" via the HP-IB removes the Range Hold
function
.
Ra-.
.....
WIttI
.....
1IaWI_
.....
......
Rnp
Hlid
1 ±
1.997Y
.OOOY
.002V
.ooov
2 ±
6.00V
2.00Y
.OlV
.OOV
Symmetry
Variable
symmetry
is
not
allowed in frequencv range 8 and conversely, frequen-
cies
2:
2MHz
can only be programmed when SYM equals
50%
and the SYM light
is
off.
"EO
1"
will
be displayed
when
these
two
parameters are in
conflict.
53